Background
Hepatitis C virus (HCV) genotyping is important for treatment and epidemiological purposes. The
Conclusion
HCV genotyping with AmpliSens(®) HCV-1/2/3-FRT using real-time PCR method provides a much simpler and more feasible workflow with shorter time compared to sequencing method. There was good concordance compared to sequencing method. However, more evaluation studies would be required to show statistical significance, and to troubleshoot discordant results. AmpliSens(®) HCV-1/2/3-FRT does differentiate between genotype but not until subtype level.
Methods
A total of 17 samples collected from December 2009 to January 2011 were analyzed. Reverse transcriptase polymerase chain reaction (PCR) was performed, followed by sequencing technique.
Results
From 17 samples, four were untypeable by AmpliSens(®) HCV-1/2/3-FRT. Eleven of 13 (84.6%) results showed concordant genotypes. A specimen that was determined as genotype 3a by sequencing was genotype 1 by the AmpliSens(®) HCV-1/2/3-FRT. Another specimen that was genotype 1 by sequencing was identified as genotype 3 by AmpliSens(®) HCV-1/2/3-FRT.