Split-Window OCT biometry in pseudophakic eyes

人工晶状体眼的分窗 OCT 生物测量

阅读:23
作者:Bartosz L Sikorski, Kenneth J Hoffer

Conclusion

The study shows small, non-significant differences between the biometric measurements obtained with REVO NX SW-OCT and IOLMaster 700 SS-OCT in pseudophakic eyes. However, SW-OCT offered significantly lower ACD and LT measurement failure rates. With high-resolution imaging, SW-OCT enables accurate assessment of IOL position relative to the posterior capsule and visualization of capsular fibrosis.

Methods

Sixty-nine pseudophakic eyes of 69 subjects were enrolled in the study. The

Purpose

To determine the utility of Split-Window optical coherence tomography OCT (SW-OCT) biometry in measuring ocular axial dimensions as well as imaging the intraocular lens (IOL) and posterior capsule in pseudophakic eyes.

Results

The correlation between measurements obtained with SW-OCT and SS-OCT was very high (ICC for: axial length (AL) = 1.000; anterior chamber depth (ACD) = 0.997; IOL thickness (IOL LT) = 0.997; central corneal thickness (CCT) = 0.987). The mean AL measurement difference was 0.003 ± 0.021 mm (the 95% LoA ranged from -0.04 to 0.05); the mean ACD difference was -0.009 ± 0.025 mm (95% LoA, -0.06 to 0.04); mean LT difference was 0.001 ± 0.021 mm (95% LoA, -0.04 to 0.04); and mean CCT difference was 1.4 ± 5.4 μm (95% LoA, -9 to 12).