Macro- and microscale variables regulate stent haemodynamics, fibrin deposition and thrombomodulin expression

宏观和微观变量调节支架血流动力学、纤维蛋白沉积和血栓调节蛋白表达

阅读:6
作者:Juan M Jiménez, Varesh Prasad, Michael D Yu, Christopher P Kampmeyer, Abdul-Hadi Kaakour, Pei-Jiang Wang, Sean F Maloney, Nathan Wright, Ian Johnston, Yi-Zhou Jiang, Peter F Davies

Abstract

Drug eluting stents are associated with late stent thrombosis (LST), delayed healing and prolonged exposure of stent struts to blood flow. Using macroscale disturbed and undisturbed fluid flow waveforms, we numerically and experimentally determined the effects of microscale model strut geometries upon the generation of prothrombotic conditions that are mediated by flow perturbations. Rectangular cross-sectional stent strut geometries of varying heights and corresponding streamlined versions were studied in the presence of disturbed and undisturbed bulk fluid flow. Numerical simulations and particle flow visualization experiments demonstrated that the interaction of bulk fluid flow and stent struts regulated the generation, size and dynamics of the peristrut flow recirculation zones. In the absence of endothelial cells, deposition of thrombin-generated fibrin occurred primarily in the recirculation zones. When endothelium was present, peristrut expression of anticoagulant thrombomodulin (TM) was dependent on strut height and geometry. Thinner and streamlined strut geometries reduced peristrut flow recirculation zones decreasing prothrombotic fibrin deposition and increasing endothelial anticoagulant TM expression. The studies define physical and functional consequences of macro- and microscale variables that relate to thrombogenicity associated with the most current stent designs, and particularly to LST.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。