Optilene, a new non-absorbable monofilament is safe and effective for CABG anastomosis. OPTICABG - A prospective international, multi-centric, cohort study

Optilene 是一种新型不可吸收单丝,可安全有效地用于 CABG 吻合术。OPTICABG - 一项前瞻性的国际、多中心、队列研究

阅读:13
作者:Adrian Ursulescu, Petra Baumann, Manel Tauron Ferrer, Monica Contino, Claudia Romagnoni, Carlo Antona, Josep Maria Padró Fernández

Conclusion

Optilene suture material represents a safe and effective alternative to existing sutures used in CABG surgery for anastomosis construction.

Methods

The study was designed as an international, multi-centre, prospective cohort study to evaluate the safety and efficacy of a new non-absorbable monofilament for CABG surgery compared to data published in a previous meta-analysis. Optilene suture was used to create the distal and proximal coronary artery anastomoses. The primary endpoint was the cumulative MACCE rate up to discharge. Secondary parameters were intraoperative handling of the suture material and QoL up to 3 months after surgery. Patients were examined 30 days and 3 months postoperatively.

Results

In total, 199 patients were enrolled in 3 centres in Europe. The cumulative CABG adverse event rate up to the day of discharge was 3%, in contrast to the 8.46% given by the data generated by Nalysnyk et al. A t-test showed that our CABG rate was significantly lower. QoL significantly increased from preoperatively until 3 months after surgery. Ease of handling the suture material was rated as very good.