The dosimetric effect of intrafraction prostate motion on step-and-shoot intensity-modulated radiation therapy plans: magnitude, correlation with motion parameters, and comparison with helical tomotherapy plans

分次治疗过程中前列腺运动对步进式调强放射治疗计划的剂量学影响:幅度、与运动参数的相关性以及与螺旋断层放射治疗计划的比较

阅读:1

Abstract

PURPOSE: To determine the daily and cumulative dosimetric effects of intrafraction prostate motion on step-and-shoot (SNS) intensity-modulated radiation therapy (IMRT) plans, to evaluate the correlation of dosimetric effect with motion-based metrics, and to compare on a fraction-by-fraction basis the dosimetric effect induced in SNS and helical tomotherapy plans. METHODS AND MATERIALS: Intrafraction prostate motion data from 486 fractions and 15 patients were available. A motion-encoded dose calculation technique was used to determine the variation of the clinical target volume (CTV) D(95%) values with respect to the static plan for SNS plans. The motion data were analyzed separately, and the correlation coefficients between various motion-based metrics and the dosimetric effect were determined. The dosimetric impact was compared with that incurred during another IMRT technique to assess correlation across different delivery techniques. RESULTS: The mean (±1 standard deviation [SD]) change in D(95%) in the CTV over all 486 fractions was 0.2 ± 0.5%. After the delivery of five and 12 fractions, the mean (±1 SD) changes over the 15 patients in CTV D(95%) were 0.0 ± 0.2% and 0.1 ± 0.2%, respectively. The correlation coefficients between the CTV D(95%) changes and the evaluated motion metrics were, in general, poor and ranged from r = -0.2 to r = -0.39. Dosimetric effects introduced by identical motion in SNS and helical tomotherapy IMRT techniques were poorly correlated with a correlation coefficient of r = 0.32 for the CTV. CONCLUSIONS: The dosimetric impact of intrafraction prostate motion on the CTV is, in general, small. In only 4% of all fractions did the dosimetric consequence exceed 1% in the CTV. As expected, the cumulative effect was further reduced with fractionation. The poor correlations between the calculated motion parameters and the subsequent dosimetric effect implies that motion-based thresholds are of limited value in predicting the dosimetric impact of intrafraction motion. The dosimetric effects between the two evaluated delivery techniques were poorly correlated.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。