Reliability and Agreement of an Integrated Platform for Intelligent Visual Function Measurement

智能视觉功能测量集成平台的可靠性和一致性

阅读:1

Abstract

INTRODUCTION: Phoropters are widely accepted for clinical use in refraction examination and visual function assessment. This study assessed the reliability of the new Inspection Platform of Visual Function (IPVF) in comparison with the conventional equipment phoropter (TOPCON VT-10) in visual function assessment. METHODS: This prospective study enrolled 80 eyes of 80 healthy subjects. The horizontal phoria at distance and near (Phoria_D and Phoria_N, respectively) was measured with the von Graefe method, negative/positive relative accommodation (NRA/PRA) was measured with the positive/negative lens method, and accommodative amplitude (AMP) was measured with the minus lens method. Data of three consecutive measurements with each instrument were evaluated using the intraclass correlation coefficient (ICC) for repeatability, and the agreement of the two instruments was evaluated using a Bland-Altman plot. RESULTS: The ICCs of the three consecutive measurements for phoria, NRA/PRA, and AMP using the IPVF instrument were high (0.87-0.96), indicating high repeatability. The ICCs of the three consecutive measurements using the phoropter were high (0.914-0.983) for phoria, NRA, and AMP, indicating high repeatability, while that of PRA was 0.732 (between 0.4 and 0.75), indicating acceptable repeatability. The 95% limits of agreement of phoria, NRA/PRA, and AMP were narrow, indicating good agreement between the two instruments. CONCLUSION: The repeatability of both instruments was high, and the IPVF instrument was slightly better in terms of PRA repeatability than the phoropter. The agreement of phoria, NRA/PRA, and AMP measured by the new IPVF instrument and phoropter was also satisfactory.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。