Intraocular Pressure Measurement with Pneumatonometry and a Tonometer Tip Cover

使用气动眼压计和眼压计探头盖进行眼内压测量

阅读:1

Abstract

PURPOSE: To investigate the precision and accuracy of IOP measurements using a pneumatonometer and a tonometer tip cover (Tono-Pen(®) tip cover) acting as a membrane between a cadaver eye model and pneumatonometer probe. METHODS: A total of 480 paired IOP measurements, with and without a Tono-Pen cover, were collected across 4 pressure levels of 7, 10, 20 and 30 mmHg. IOP measurements were obtained by three different pneumatonometer units paired with three different masked operators (three configurations). Four eyes were sampled for each eye pressure level. The sequence of eye pressure, configuration, and measurements with vs. without the Tono-Pen cover was randomized. RESULTS: With the Tono-Pen cover in place, there was a negative bias with a mean IOP difference of - 1.18 mmHg for all 480 paired samples compared with the measurements absent the cover. Compared with the test pressure settings (i.e., 7, 10, 20, 30 mmHg), the overall mean bias was + 0.35 mmHg with the Tono-Pen cover present. With the Tono-Pen cover present, the overall repeatability %CV (percent coefficient of variation) was 3.4% and the reproducibility %CV was 3.8% compared with a repeatability %CV of 3.2% and reproducibility %CV of 5.7% without the Tono-Pen cover. CONCLUSION: Measurement of IOP via pneumatonometry with a Tono-Pen cover in place, also known as the excursion test method, yields precise, accurate and reproducible results. This developed method of pressure measurement is an acceptable and reliable form of IOP measurement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。