Abstract

Recently, computed tomography with photon-counting detector (PCD-CT) has been developed to enable high-resolution imaging at a lower radiation dose. PCD-CT employs a photon-counting detector that can measure the number of incident X-ray photons and their energy. The newly released PCD-CT (NAEOTOM Alpha, Siemens Healthineers, Forchheim, Germany) has been in clinical use at our institution since December 2022. The PCD-CT offers several advantages over current state-of-the-art energy-integrating detector CT (EID-CT). The PCD-CT does not require septa to create a detector channel, while EID-CT does. Therefore, downsizing the anode to achieve higher resolution does not affect the dose efficiency of the PCD-CT. CT is an indispensable modality for evaluating ear ossicles. The ear ossicles and joints are clearly depicted by PCD-CT. In particular, the anterior and posterior legs of the stapes, which are sometimes unclear on conventional CT scans, can be clearly visualized. We present cases of congenital anomalies of the ossicular chain, ossicular chain dislocation, tympanosclerosis, and cholesteatoma in which PCD-CT was useful. This short article reports the usefulness of PCD-CT in the 3D visualization of the ear ossicles.