Precision of volumetric assessment of proximal femur microarchitecture from high-resolution 3T MRI

利用高分辨率3T磁共振成像技术对近端股骨微结构进行体积评估的精确性

阅读:1

Abstract

PURPOSE: To evaluate the precision of measures of bone volume and bone volume fraction derived from high-resolution 3T MRI of proximal femur bone microarchitecture using non-uniformity correction. METHODS: This HIPAA compliant, institutional review board approved study was conducted on six volunteers (mean age 56 ± 13 years), and written informed consent was obtained. All volunteers underwent a 3T FLASH MRI hip scan at three time points: baseline, second scan same day (intra-scans), and third scan one week later (inter-scans). Segmentation of femur images and values for total proximal femur volume (T), bone volume (B), and bone volume fraction (BVF) were calculated using in-house developed software, FireVoxel. Two types of non-uniformity corrections were applied to images (N3 and BiCal). Precision values were calculated using absolute percent error (APE). Statistical analysis was carried out using one-sample one-sided t test to observe the consistency of the precision and paired t test to compare between the various methods and scans. RESULTS: No significant differences in bone volume measurements were observed for intra- and inter-scans. When using non-uniformity correction and assessing all subjects uniformly at the level of the lesser trochanter, precision values overall improved, especially significantly (p < 0.05) when measuring bone volume, B . B values using the combination of N3 or BiCal with CLT had a significant consistent APE values of less than 2.5 %, while BVF values were all consistently and significantly lower than 2.5 % APE. CONCLUSION: Our results demonstrate the precision of high-resolution 3D MRI measures were comparable to that of dual-energy X-ray absorptiometry. Additional corrections to the analysis technique by cropping at the lesser trochanter or using non-uniformity corrections helped to improve precision. The high precision values from these MRI scans provide evidence for MRI of the proximal femur as a promising tool for osteoporosis diagnosis and treatment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。