Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies

人工心脏瓣膜流体动力学测试中收缩期和舒张期的持续时间:ISO 5840 标准与体内研究的比较

阅读:1

Abstract

OBJECTIVE: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review. METHODS: The systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves. RESULTS: Based on the in vivo studies analyzed, the systolic durations proposed by the ISO 5840 standard seemed consistent for 45 and 120 beats per minute (bpm), and showed diverse results for the 70 bpm condition. CONCLUSION: Information on the realistic validation of the operation of left ventricular models for different heart rates were obtained.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。