Heating of Large Endovascular Stents and Stent Grafts in Magnetic Particle Imaging-Influence of Measurement Parameters and Isocenter Distance

磁粒子成像中大型血管内支架和覆膜支架的加热——测量参数和等中心距离的影响

阅读:3

Abstract

PURPOSE: Magnetic particle imaging (MPI) is a tomographic imaging modality with the potential for cardiovascular applications. In this context, the extent to which stents are heated should be estimated from safety perspective. Furthermore, the influence of the measurement parameters and stent distance to the isocenter of the MPI scanner on stent heating were evaluated. MATERIALS AND METHODS: Nine different endovascular stents and stent grafts were tested in polyvinyl-chloride tubes. The stents had diameters from 10 to 31 mm, lengths between 25 and 100 mm and were made from stainless steel, nitinol or cobalt-chromium. The temperature differences were recorded with fiber-optic thermometers. All measurements were performed in a preclinical commercial MPI scanner. The measurement parameters were varied (drive field strengths: 3, 6, 9, 12 mT and selection field gradients: 0, 1.25 and 2.5 T/m). Furthermore, measurements with different distances to the scanner's isocenter were performed (100 to 0 mm). RESULTS: All stents showed heating (maximum 53.1 K, minimum 4.6 K). The stent diameter directly correlated with the temperature increase. The drive field strength influenced the heating of the stents, whereas the selection field gradient had no detectable impact. The heating of the stents decreased with increasing distance from the scanner's isocenter and thus correlated with the loss of the scanner's magnetic field. CONCLUSION: Stents can cause potentially harmful heating in MPI. In addition to the stent diameter and design, the drive field strength and the distance to the MPI scanner's isocenter must be kept in mind as influencing parameters.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。