Thermal Behavior of Graphene Oxide Deposited on 3D-Printed Polylactic Acid for Photothermal Therapy: An Experimental-Numerical Analysis

用于光热疗法的3D打印聚乳酸上沉积氧化石墨烯的热行为:实验-数值分析

阅读:1

Abstract

The present work evaluates the thermal behavior of graphene oxide (GO) when deposited on 3D-printed polylactic acid (PLA), in order to develop a medical device for photothermal therapy applications. An experimental-numerical analysis was performed to assess the photothermal conversion capacity, based on the power emitted by a NIR (785 nm) laser, and the subsequent temperature distribution on the GO-PLA material. The influence of the deposited mass of GO and the PLA thickness was studied through 40 different scenarios. The results estimated a value of photothermal conversion efficiency of up to 32.6%, achieved for the lower laser power density that was tested (0.335 mW/mm²), and a high mass value of deposited GO (1.024 × 10(-3) mg/mm²). In fact, an optimal mass of GO in the range of 1.024-2.048 × 10(-3) mg/mm(2) is proposed, in terms of absorption capacity, since a higher mass of GO would not increase the conversion efficiency. Moreover, the study allowed for an estimation of the thermal conductivity of this specific biomaterial (0.064 W/m·K), and proved that a proper combination of GO mass, PLA thickness, and laser power can induce ablative (>60 °C, in a concentrated area), moderate (50 °C), and mild (43 °C) hyperthermia on the bottom face of the biomaterial.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。