Immediate and six months clinical and angiographic results of intracoronary paclitaxel-coated stent implantation - the Meo:DrugStar-1 study

Meo:DrugStar-1 研究:冠状动脉内紫杉醇涂层支架植入术的即刻及六个月临床和血管造影结果

阅读:1

Abstract

The study was conducted to evaluate the clinical and angiographic results of the implantation of the paclitaxel-eluting stent Meo:DrugStar ST in patients with symptomatic coronary artery disease. The Meo:DrugStar ST stent has a stainless steel stent platform with a homogenous non-biodegradable coating of paclitaxel mixed with a polyether-based biostable, monophase, and hemocompatible coating. Sixty patients with native coronary artery disease were included in the study. The Meo:DrugStar ST stents were implanted in 60 de novo lesions detected in these patients. Immediate and long-term clinical and angiographic follow-up results were evaluated. There was a high proportion of patients with hypertension (55%) according to JNC-VII. Mean stenosis ratio was 78 +/- 13 %, mean implanted stent diameter was 3.0 +/- 0.4 mm and mean length was 22 +/- 5 mm. Restenosis was detected in 4 (10%) of those patients and 11 (27.5%) of 40 patients had insignificant amount of restenosis. The results of this study indicate a potential benefit of the Meo:DrugStar ST stent for the prevention of stent thrombosis and restenosis in these relatively high-risk patients.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。