Multimodal ultrasound tomography for breast imaging: a prospective study of clinical feasibility

多模态超声断层扫描在乳腺成像中的应用:一项前瞻性临床可行性研究

阅读:1

Abstract

BACKGROUND: To describe the clinical set-up and evaluate the feasibility of multimodal ultrasound tomography (MUT) for breast imaging. METHODS: Thirty-two consecutive patients referred for breast imaging and 24 healthy volunteers underwent MUT. In the 32 patients, the examination discomfort was compared to that of mammography (n = 31), handheld ultrasound (HUS) (n = 27) and magnetic resonance imaging (MRI) (n = 4) on a scale from 1 (lowest discomfort) to 10 (highest discomfort). MUT investigation time was recorded. Findings automatically detected by MUT were correlated with conventional imaging and biopsy results. RESULTS: Breast MUT was well tolerated by all 56 participants; 55 bilateral exams were uneventful. During one exam, the digitalisation card failed and the exam was successfully repeated within three days. Mean examination discomfort was 1.6 (range = 1-5) for MUT, 1.5 (range = 1-5) for HUS, 5.3 (range = 3-7) for MRI, and 6.3 (range = 1-10) for mammography. MUT examination time was 38 ± 6 min (mean ± standard deviation). In the patients referred for breast imaging, MUT detected four lesions and indicated malignancy in three of these cases. These findings were confirmed by additional imaging and biopsy. CONCLUSION: MUT is feasible in a clinical context considering examination time and patient acceptance. These interesting initial diagnostic findings warrant further studies.