Acute Hemodynamic Effects of Cardiac Resynchronization Therapy Versus Alternative Pacing Strategies in Patients With Left Ventricular Assist Devices

左心室辅助装置患者接受心脏再同步治疗与其他起搏策略的急性血流动力学效应比较

阅读:1

Abstract

Background The hemodynamic effects of cardiac resynchronization therapy in patients with left ventricular assist devices (LVADs) are uncharacterized. We aimed to quantify the hemodynamic effects of different ventricular pacing configurations in patients with LVADs, focusing on short-term changes in load-independent right ventricular (RV) contractility. Methods and Results Patients with LVADs underwent right heart catheterization during spontaneous respiration without sedation and with pressures recorded at end expiration. Right heart catheterization was performed at different pacemaker configurations (biventricular pacing, left ventricular pacing, RV pacing, and unpaced conduction) in a randomly generated sequence with >3 minutes between configuration change and hemodynamic assessment. The right heart catheterization operator was blinded to the sequence. RV maximal change in pressure over time normalized to instantaneous pressure was calculated from digitized hemodynamic waveforms, consistent with a previously validated protocol. Fifteen patients with LVADs who were in sinus rhythm were included. Load-independent RV contractility, as assessed by RV maximal change in pressure over time normalized to instantaneous pressure, was higher in biventricular pacing compared with unpaced conduction (15.7±7.6 versus 11.0±4.0 s(-1); P=0.003). Thermodilution cardiac output was higher in biventricular pacing compared with unpaced conduction (4.48±0.7 versus 4.38±0.8 L/min; P=0.05). There were no significant differences in heart rate, ventricular filling pressures, or atrioventricular valvular regurgitation across all pacing configurations. Conclusions Biventricular pacing acutely improves load-independent RV contractility in patients with LVADs. Even in these patients with mechanical left ventricular unloading via LVAD who were relative pacing nonresponders (required LVAD support despite cardiac resynchronization therapy), biventricular pacing was acutely beneficial to RV contractility.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。