Comparison of the accuracy of full head cone beam CT images obtained using a large field of view and stitched images

比较使用大视野和拼接图像获得的全头部锥形束CT图像的准确性

阅读:1

Abstract

OBJECTIVE: The aim of this study was to investigate the accuracy of the new stitching feature of the Morita 3D Accuitomo device by comparing it with the I-CAT device and real measurements. METHODS: Eight dry skulls were used in this study. The full scan mode of an I-CAT device and the new stitching feature of the Morita 3D Accuitomo device were used to obtain full head cone beam CT (CBCT) images, and the linear distances between the selected anatomical sites were measured by two observers. The real distances were measured directly on the skulls using digital calipers. Statistical analyses were performed using the intraclass correlation coefficient and Bland-Altman Limits of Agreement. RESULTS: High or perfect agreement was detected between the CBCT images and real measurements. The Bland-Altman Limits of Agreement analysis revealed that the stitched coronal and sagittal images of the Morita might differ by >1.54-1.67 or <0.01 mm from the real measurements. CONCLUSIONS: The stitched images of the Morita CBCT device showed perfect agreement with both real measurements and the images of an I-CAT full head scan. After the evaluation of the radiation doses received from stitching sectional images, this device may be used to obtain full head images.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。