Added value of a connected glucose meter for glycorrhachia assessment

联网血糖仪在血糖评估中的附加价值

阅读:2

Abstract

OBJECTIVES: The aim of this study was to demonstrate the performance and added value of rapid glucose determination in cerebrospinal fluid using a connected glucometer. DESIGN AND METHODS: Intra-assay and inter-assay accuracies were calculated using residual clinical samples. Accuracies were measured by comparing the results obtained with the glucometer to those from the central laboratory on a large routine chemistry platform. RESULTS: The intra-assay coefficients of variation were between 6.1% and 6.2% for low values (18 mg/dL) and between 5.6% and 6.8% for high values (58 mg/dL). The inter-assay coefficients of variation were between 9.4% and 16.3% for the low values (18 mg/dL) and between 5.7% and 8.7% for the high values (pool; ±75 mg/dL). The regression equation by comparison to the central laboratory was y = 4.08 + 0.82 x, with a coefficient of determination (r(2)) of 0.95. CONCLUSIONS: The measurement of glycorrhachia with a connected glucometer before the analysis in the central laboratory allows a rapid orientation in the deferential diagnosis of a meningitis of viral vs bacterial origin. The response time is fast (6 s) and requires only a small amount of fluid (1.2 μL), which is important in infants, especially since lumbar puncture is an integral part of the investigation of the origin of a fever in this population.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。