Ingestible sensors correlate closely with peripheral temperature measurements in febrile patients

可吞咽式传感器与发热患者的外周体温测量结果密切相关

阅读:1

Abstract

BACKGROUNDS: Reliable non-invasive methods for measuring body temperature are essential for the diagnosis and monitoring of infectious disease. METHODS: This study used Intraclass Correlation Coefficients (ICC) and the Bland- Altman plot to analyse the agreement between temperature measurements using an ingestible capsule sensor, a skin sensor and two non-invasive peripheral temperature measurements (axillary and infrared non-contact), collected from a population of febrile patient admitted for infectious disease. RESULTS: Of the 77 febrile patients screened, 26 patients were enrolled. The ICC between axillary temperature measurements (Taxi) vs. non-contact measurements (Tno-c) were 0.34 [-0.18; 0.63], 0.87 [0.55; 0.94] between Taxi vs. ingestible capsule measurements (Tcap) and 0.12 [-0.09; 0.37] between Taxi vs. Tetac. The mean difference between Taxi vs Tno-c was -1.18 °C with limits of agreement (LoA) from -2.96 to 0.58 °C. The mean difference between Taxi vs Tcap was 0.48 °C, with LoA from -0.60 to 1.56 °C. The mean difference between Taxi vs Tetac was -4.23 °C with LoA from -7.22 to -1.23 °C. CONCLUSIONS: Ingestible capsule measurements are reliable enough to adequately estimate the core body temperature in clinical practice. Its non-invasiveness, and the real-time remote control offer new opportunities for future research into fever during infectious diseases.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。