Implementation of optically simulated luminescent dosimeter for quality control of gamma ray dose of an accelerator-based neutron source

利用光学模拟发光剂量计对加速器中子源的伽马射线剂量进行质量控制

阅读:1

Abstract

BACKGROUND: Neutron beams utilized for performing BNCT are composed of a mixture of neutrons and gamma rays. Although much of the dose delivered to the cancer cells comes from the high LET particles produced by the boron neutron capture reaction, the dose delivered to the healthy tissues from unwanted gamma rays cannot be ignored. With the increase in the number of accelerators for BNCT, a detector system that is capable of measuring gamma ray dose in a mixed neutron/gamma irradiation field is crucial. Currently, BeO TLDs encased in quartz glass are used to measure gamma ray dose in a BNCT irradiation field. However, this type of TLD is no longer commercially available. A replacement dosimetry system is required to perform the recommended ongoing quality assurance of gamma ray measurement for a clinical BNCT system. PURPOSE: The purpose of this study is to evaluate the characteristics of a BeO OSLD detector system under a mixed neutron and gamma ray irradiation field and to assess the suitability of the system for routine quality assurance measurements of an accelerator-based BNCT facility. METHODS: The myOSLD system by RadPro International GmbH was evaluated using the accelerator-based neutron source designed for clinical BNCT (NeuCure BNCT system). The readout constancy, linearity, dose rate effect, and fading effect of the OSLD were evaluated. Free-in-air and water phantom measurements were performed and compared with the TLD results and Monte Carlo simulation results. The PHITS Monte Carlo code was used for this study. RESULTS: The readout constancy was found to be stable over a month-long period and similar to the TLD results. The OSLD readout signal was found to be linear, with a high coefficient of determination (R(2) ≥ 0.999) up to a proton charge of 3.6 C. There was no significant signal fading or dose rate dependency. The central axis depth dose and off-axis dose profile measurements agreed with both the TLD and Monte Carlo simulation results, within one standard deviation. CONCLUSION: The myOSLD system was characterized using an accelerator system designed for clinical BNCT. The experimental measurements confirmed the OSLD achieved similar, if not superior to, the currently utilized dosimetry system for routine QA of an accelerator-based BNCT system. The OSLD system would be a suitable replacement for the current TLD system for performing routine QA of gamma ray dose measurement in a BNCT irradiation field.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。