The delivery assessment for small targets on Halcyon radiotherapy system - Measured and calculated dose comparison

Halcyon放射治疗系统小靶区剂量输送评估——实测剂量与计算剂量对比

阅读:1

Abstract

BACKGROUND: With the ever-increasing requirements of accuracy and personalization of radiotherapy treatments, stereotactic radiotherapy (SRT) with volumetric modulated arc therapy (VMAT) on O-ring Halcyon radiotherapy system could potentially provide a fast, safe, and feasible treatment option. PURPOSE: The purpose of this study was to assess the delivery of Halcyon VMAT plans for small targets. METHODS: Well-defined VMAT-SRT plans were created on Halcyon radiotherapy system with the stacked and staggered dual-layer MLC design for the film measurement set-up and the target sizes and shapes designed to emulate the targets of the stereotactic treatments. The planar dose distributions were acquired with film measurements and compared to a current clinical reference dose calculation with AcurosXB (v18.0, Varian Medical Systems) and to Monte Carlo simulations. With the collapsed arc versions of the VMAT-SRT plans, the uncertainty in dose delivery due to the multileaf collimator (MLC) without the gantry rotation could be separated and analyzed. RESULTS: The target size was mainly limited by the resolution originated from the design of the MLC leaves. The results of the collapsed arc versions of the plans show good consistency among measured, calculated, and simulated dose distributions. With the full VMAT plans, the agreement between calculated and simulated dose distributions was consistent with the collapsed arc versions. The measured dose distribution agreed with the calculated and simulated dose distributions within the target regions, but considerable local differences were observed in the margins of the target. The largest differences located in the steep gradient regions presumably originating from the deviation of the isocenter. CONCLUSIONS: The potential of the Halcyon radiotherapy system for VMAT-SRT delivery was evaluated and the study revealed valuable insights on the machine characteristics with the delivery.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。