Dosimetric impact of rotational errors in trigeminal neuralgia radiosurgery using CyberKnife

使用CyberKnife进行三叉神经痛放射外科手术时旋转误差的剂量学影响

阅读:1

Abstract

PURPOSE: Trigeminal neuralgia (TN) can be treated on the CyberKnife system using two different treatment delivery paths: the general-purpose full path corrects small rotations, while the dedicated trigeminal path improves dose fall-off but does not allow rotational corrections. The study evaluates the impact of uncorrected rotations on brainstem dose and the length of CN5 (denoted as L(eff)) covered by the prescription dose. METHODS AND MATERIALS: A proposed model estimates the delivered dose considering translational and rotational delivery errors for TN treatments on the CyberKnife system. The model is validated using radiochromic film measurements with and without rotational setup error for both paths. L(eff) and the brainstem dose is retrospectively assessed for 24 cases planned using the trigeminal path. For 15 cases, plans generated using both paths are compared for the target coverage and toxicity to the brainstem. RESULTS: In experimental validations, measured and estimated doses agree at 1%/1 mm level. For 24 cases, the treated L(eff) is 5.3 ± 1.7 mm, reduced from 5.9 ± 1.8 mm in the planned dose. Constraints for the brainstem are met in 23 cases for the treated dose but require frequent treatment interruption to maintain rotational corrections <0.5° using the trigeminal path. The treated length of CN5, and plan quality metrics are similar for the two paths, favoring the full path where rotations are corrected. CONCLUSIONS: We validated an analytical model that can provide patient-specific tolerances on rotations to meet plan objectives. Treatment using the full path can reduce treatment time and allow for rotational corrections.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。