The implementation of an image-guided system at a proton therapy center facility

在质子治疗中心设施中实施图像引导系统

阅读:1

Abstract

PURPOSE: Pencil Beam Scanning (PBS) proton therapy has similar requirements on patient alignment to within 1 mm and 1-degree accuracy as photon radiosurgery. This study describes general workflow, acceptance, and commissioning test procedures and their respective results for an independent robotic arm used for Image Guided Radiotherapy (IGRT) for a Proton Therapy System. METHODS: The system is equipped with kV-imaging techniques capable of orthogonal and Cone-Beam Computed Tomography (CBCT) imaging modalities mounted on an independent robotic arm gantry attached to the ceiling. The imaging system is capable of 360-degree rotation around patients to produce CBCT and kilovoltage orthogonal images. The imaging hardware is controlled by Ehmet Health XIS software, and MIM Software handles the image fusion and registration to an acceptable accuracy of ≤1-mm shifts for patients' alignment. The system was tested according to the requirements outlined in the American Association of Physicists in Medicine (AAPM) Task Group (TG) 142 and TG 179. The system tests included (1) safety, functionality, and connectivity, (2) mechanical testing, (3) image quality, (4) image registration, and (5) imaging dose. Additional tests included imaging gantry isocentricity with a laser tracker and collision-avoiding system checks. RESULTS: The orthogonal and volumetric imaging are comparable in quality to other commercially available On-Board Imagers (OBI) systems. The resulting spatial resolution values were 1.8-, 0.8-, and 0.5-Line Pairs per Millimeter (lp/mm) for orthogonal, full-fan CBCT, and half-fan CBCT, respectively. The image registration is accurate to within 1 mm and 1 degree. The data shows consistent imaging-guided system performance with standard deviations in x, y, and z of 0.7, 0.8, and 0.7 mm, respectively. CONCLUSIONS: The system provides excellent image quality and performance, which can be used for IGRT. The proven accuracy of the x-ray imaging and positioning system at McLaren Proton Therapy Center (MPTC) is 1 mm, making it suitable for proton therapy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。