Angular correction methodology and characterization of a high-resolution CMOS array for patient specific quality assurance on a robotic arm linac

用于机器人手臂直线加速器上患者特定质量保证的高分辨率CMOS阵列的角度校正方法和特性分析

阅读:1

Abstract

PURPOSE: To develop an angular correction methodology and characterize a high-resolution complementary metal-oxide-semiconductor (CMOS) array for patient specific quality assurance on a robotic arm linear accelerator. METHODS: Beam path files from the treatment planning software (TPS) were used to calculate the angle of radiation beam with respect to the detector plane. Beams from multiple discrete angles were delivered to the CMOS detector array and an angular dependency look up table (LUT) was created. The LUT was then used to correct for the angular dependency of the detector. An iso-centric 5 mm fixed cone, non iso-centric multi-target fixed cone, 10 mm Iris and a multi-leaf collimator (MLC) based collimated plan were delivered to the phantom and compared to the TPS with and without angular correction applied. Additionally, the CMOS array was compared to gafchromic film and a diode array. RESULTS: Large errors of up to 30% were observed for oblique angles. When angular correction was applied, the gamma passing rate increased from 99.2% to 100% (average gamma value decreased from 0.29 to 0.14) for the 5-mm iso-centric cone plan. Similarly, the passing rate increased from 84.0% to 100% for the Iris plan and from 49.98% to 98.4% for the MLC plan when angular correction was applied. For the multi-target plan, applying angular correction improved the gamma passing rate from 94% to 99.6%. The 5 mm iso-centric fixed cone plan was also delivered to film, and the gamma passing rate was 91.3% when using gafchromic film as the reference dataset, whereas the diode array provided insufficient sampling for this plan. CONCLUSION: A methodology of calculating the beam angle based on the beam path files was developed and validated. The array was demonstrated to be superior to other quality assurance tools because of its sub-millimeter spatial resolution and immediate read out of the results.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。