Relationship between imaging parameters and distortion in magnetic resonance images for gamma knife treatment planning

伽玛刀治疗计划中成像参数与磁共振图像畸变的关系

阅读:1

Abstract

In magnetic resonance imaging (MRI), it is necessary to reduce image distortion as much as possible because it suppresses the increase in the planning target volume. This study investigated the relationship between imaging parameters and image distortion when using G-frames. The images were obtained using a 1.5-T MRI system with a 09-101 Pro-MRI phantom. Image distortion was measured by changing the RF pulse mode, gradient mode, asymmetric echo, and bandwidth (BW). The image distortion was increased in the high RF mode than in the Normal mode. The image distortion increased in the following order: Whisper ≦ Normal < Fast in the different gradient modes. The image distortion increased in the following order: Without ≦ Weak < Strong in the different asymmetric echo modes. The image distortion increased in the following order: 300 Hz/pixel > 670 Hz/pixel ≧ REF (150 Hz/pixel) in the different Bw. The relationship between parameters and image distortion was clarified in this study when G-frames were used for gamma knife therapy. There is had relationship between the parameters causing variation in the gradient magnetic field and image distortion. Therefore, these parameters should be adjusted to minimize distortion.