Gamma passing rates of daily EPID transit images correlate to PTV coverage for breast cancer IMRT treatment plans

每日EPID过境图像的伽马通过率与乳腺癌IMRT治疗计划的PTV覆盖率相关

阅读:2

Abstract

PURPOSE: The use of the transit image obtained with the electronic portal-imaging device (EPID) is becoming an extended method to perform in-vivo dosimetry. The transit images acquired during each fraction can be compared with a predicted image, if available, or with a baseline image, usually the obtained in the first fraction. This work aims to study the dosimetric impact of the failing fractions and to evaluate the appropriateness of using a baseline image in breast plans. MATERIAL AND METHODS: Twenty breast patients treated in a Halcyon were retrospectively selected. For each patient and fraction, the treatment plan was calculated over the daily CBCT image. For each fraction, the differences respect to the treatment plan values of OARs and PTV dosimetric parameters were analyzed: ΔD(mean) , ΔD95%, ΔD98%, ΔD2%, ΔV36Gy, ΔV38.5Gy, and ΔV43.5Gy. Daily fractions were ranked according to the differences found in the dosimetric parameters between the treatment plan and the daily CBCT to establish the best fraction. The daily transit images acquired in every fraction were compared to the first fraction using the global gamma index with the Portal Dosimetry tool. The comparison was repeated using the best fraction image as a baseline. We assessed the correlation of the dosimetric differences obtained from the CBCT images-based treatment plans with the gamma index passing rates obtained using first fraction and best fraction as baseline. RESULTS: Average values of -11.6% [-21.4%, -3.3%] and -3.2% [-1.0%, -10.3%] for the ∆PTVD98% and ∆PTVD95% per every 10% decrease in the passing rate were found, respectively. When using the best fraction as baseline patients were detected with failing fractions that were not detected with the first fraction as baseline. CONCLUSION: The gamma passing rates of daily transit images correlate with the coverage loss parameters in breast IMRT plans. Using first fraction image as baseline can lead to the non-detectability of failing fractions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。