Abstract
PURPOSE: To quantify the potential error in outputs for flattening filter free (FFF) beams associated with use of a lead foil in beam quality determination per the addendum protocol for TG-51, we examined differences in measurements of the beam quality conversion factor k(Q) when using or not using lead foil. METHODS: Two FFF beams, a 6 MV FFF and a 10 MV FFF, were calibrated on eight Varian TrueBeams and two Elekta Versa HD linear accelerators (linacs) according to the TG-51 addendum protocol by using Farmer ionization chambers [TN 30013 (PTW) and SNC600c (Sun Nuclear)] with traceable absorbed dose-to-water calibrations. In determining k(Q) , the percentage depth-dose at 10 cm [PDD(10)] was measured with 10×10 cm(2) field size at 100 cm source-to-surface distance (SSD). PDD(10) values were measured either with a 1 mm lead foil positioned in the path of the beam [%dd(10)(Pb) ] or with omission of a lead foil [%dd(10)]. The %dd(10)x values were then calculated and the k(Q) factors determined by using the empirical fit equation in the TG-51 addendum for the PTW 30013 chambers. A similar equation was used to calculate k(Q) for the SNC600c chamber, with the fitting parameters taken from a very recent Monte Carlo study. The differences in k(Q) factors were compared for with lead foil vs. without lead foil. RESULTS: Differences in %dd(10)x with lead foil and with omission of lead foil were 0.9 ± 0.2% for the 6 MV FFF beam and 0.6 ± 0.1% for the 10 MV FFF beam. Differences in k(Q) values with lead foil and with omission of lead foil were -0.1 ± 0.02% for the 6 MV FFF and -0.1 ± 0.01% for the 10 MV FFF beams. CONCLUSION: With evaluation of the lead foil role in determination of the k(Q) factor for FFF beams. Our results suggest that the omission of lead foil introduces approximately 0.1% of error for reference dosimetry of FFF beams on both TrueBeam and Versa platforms.