Evaluation of deep-learning image reconstruction for chest CT examinations at two different dose levels

评估深度学习图像重建在两种不同剂量水平下对胸部CT检查的应用

阅读:1

Abstract

AIMS: The aims of the present study were to, for both a full-dose protocol and an ultra-low dose (ULD) protocol, compare the image quality of chest CT examinations reconstructed using TrueFidelity (Standard kernel) with corresponding examinations reconstructed using ASIR-V (Lung kernel) and to evaluate if post-processing using an edge-enhancement filter affects the noise level, spatial resolution and subjective image quality of clinical images reconstructed using TrueFidelity. METHODS: A total of 25 patients were examined with both a full-dose protocol and an ULD protocol using a GE Revolution APEX CT system (GE Healthcare, Milwaukee, USA). Three different reconstructions were included in the study: ASIR-V 40%, DLIR-H, and DLIR-H with additional post-processing using an edge-enhancement filter (DLIR-H + E2). Five observers assessed image quality in two separate visual grading characteristics (VGC) studies. The results from the studies were statistically analyzed using VGC Analyzer. Quantitative evaluations were based on determination of two-dimensional power spectrum (PS), contrast-to-noise ratio (CNR), and spatial resolution in the reconstructed patient images. RESULTS: For both protocols, examinations reconstructed using TrueFidelity were statistically rated equal to or significantly higher than examinations reconstructed using ASIR-V 40%, but the ULD protocol benefitted more from TrueFidelity. In general, no differences in observer ratings were found between DLIR-H and DLIR-H + E2. For the three investigated image reconstruction methods, ASIR-V 40% showed highest noise and spatial resolution and DLIR-H the lowest, while the CNR was highest in DLIR-H and lowest in ASIR-V 40%. CONCLUSION: The use of TrueFidelity for image reconstruction resulted in higher ratings on subjective image quality than ASIR-V 40%. The benefit of using TrueFidelity was larger for the ULD protocol than for the full-dose protocol. Post-processing of the TrueFidelity images using an edge-enhancement filter resulted in higher image noise and spatial resolution but did not affect the subjective image quality.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。