Determination of correction factors in small MLC-defined fields for the Razor and microSilicon diode detectors and evaluation of the suitability of the IAEA TRS-483 protocol for multiple detectors

确定 Razor 和微硅二极管探测器在小 MLC 定义视野中的校正因子,并评估 IAEA TRS-483 协议对多个探测器的适用性

阅读:1

Abstract

Small field output factors for Multileaf collimator (MLC)-defined field sizes between 0.5 × 0.5 cm(2) and 3 × 3 cm(2) were measured with six different detectors for a Varian TrueBeam in 6-MV, 6-FFF, 10-MV, and 10-FFF photon beams. Correction factors kQclin,Qreffclin,fref from the IAEA publication TRS-483 were used to correct the measured output factors. The corrected output factors from the six detectors were used to calculate correction factors for the PTW microSilicon T60023 (PTW, Freiburg, Germany) and IBA Razor (IBA Dosimetry, Schwarzenbruck, Germany) detectors. The uncertainty of the output and correction factors in this study was calculated and the calculations presented in detail. The application of the TRS-483 correction factors significantly reduced the variation in output factors between the various detectors, with the exception of the PTW 60016 diode in 6-MV and 6-FFF beams, and the IBA PFD in 10-MV and 10-FFF beams. Correction factors calculated for the Razor agreed within 2.9% of existing literature for all energies, while the microSilicon correction factors agreed within 1.6% to the literature for 6-MV beams. The uncertainty in the microSilicon and Razor correction factors was calculated to be less than 0.9% (k = 1). This study shows that TRS-483 correction factors reduce the variation in output factors between the detectors used in this study and presents a suitable method for determining correction factors for detectors with unpublished values.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。