Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy

实施并评估高分辨率二极管阵列,用于机器人脑立体定向放射外科/放射治疗的患者个体化质量保证

阅读:1

Abstract

The purpose of the study was to introduce and evaluate a high-resolution diode array for patient-specific quality assurance (PSQA) of CyberKnife brain stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT). Thirty-three intracranial plans were retrospectively delivered on the SRS MapCHECK using fixed cone, Iris, and multileaf collimator (MLC). The plans were selected to cover a range of sites from large tumor bed, single/multiple small brain metastases (METs) to trigeminal neuralgia. Fiducial tracking using the four fiducials embedded around the detector plane was used as image guidance. Results were analyzed before and after registration based on absolute dose gamma criterion of 1 mm distance-to-agreement and 0.5%-3% dose-difference. Overall, the gamma passing rates (1 mm and 3% criterion) before registration for all the patients were above 90% for all three treatment modalities (96.8 ± 3.5%, the lowest passing rate of 90.4%), and were improved after registration (99.3 ± 1.5%). When tighter criteria (1 mm and 2%) were applied, the gamma passing rates after registration for all the cases dropped to 97.3 ± 3.2%. For trigeminal neuralgia cases, we applied 1 mm and 0.5% criterion and the passing rates dropped from 100 ± 0.0% to 98.5 ± 2.0%. The mean delivery time was 33.4 ± 11.7 min, 24.0 ± 4.9 min, and 17.1 ± 2.6 min for the fixed cone, Iris, and MLC, respectively. With superior gamma passing rates and reasonable quality assurance (QA) time, we believe the SRS MapCHECK could be a good option for routine PSQA for CyberKnife SRS/SRT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。