A measure of SRS/SRT plan quality: Quantitative limits for intermediate dose spill (R50%) in linac-based delivery

立体定向放射外科/立体定向放射治疗计划质量的衡量标准:基于直线加速器的治疗中中间剂量溢出(R50%)的定量限值

阅读:1

Abstract

Stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT) of multiple cranial targets using a single isocenter on conventional C-arm linear accelerators are rapidly developing clinical techniques. However, no universal guidelines for acceptable intermediate dose spill limits are currently available or widely accepted. In this work, we propose an intermediate dose spill guidance range for cranial SRS/SRT delivered on C-arm linacs with MLC collimation for single PTV plans and single isocenter multiple target plans with PTV volumes in the range 0.02-57.9 cm(3) . We quantify intermediate dose spill with the R50% metric (R50% = volume of 50% of prescription isodose cloud / volume of PTV) and test the proposed range using three clinical data sets, containing both 6 MV and 10 MV beams, previously published by other authors. Our proposed lower limit of R50% (LowerR50%) and upper limit of acceptable R50% (UpperR50%) bound over 90% of the clinical data used in this study, yet still provide a challenging benchmark for optimization and plan assessment of linac-based, MLC collimated SRS/SRT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。