Technical note: On the impact of the kV imaging configuration on doses from planar images during motion-synchronized treatments on Radixact®

技术说明:关于kV成像配置对Radixact®运动同步治疗期间平面图像剂量的影响

阅读:1

Abstract

Kilovoltage radiographs are acquired during motion-synchronized treatments on Radixact to localize the tumor during the treatment. Several previous publications have provided estimates of patient dose from these planar radiographs. However, a recent hardware update changed several aspects of the kV imaging system, including a new X-ray tube, an extended source-to-axis distance (SAD), and a larger field size. This is denoted the extended configuration. The purpose of this work was to assess the impact of the configuration change on patient dose from these procedures. Point doses in water were measured using the TG-61 protocol for tube potentials between 100 and 140 kVp for both the standard and extended configurations under the same water tank setup. Comparisons were made for equal mAs since the same protocols (kVp, mAs) will be used for both configurations. In comparison to the standard configuration, doses per mAs from the extended configuration were found to be ~66% less and falloff less steep due to the increased SAD. However, a larger volume of tissue is irradiated due to the larger field size. Beam quality for a given tube potential was the same as determined by half-value layer measurements. Both kV configurations are available from the vendor, therefore, the values in this work can be used to compare values previously published in the literature for the standard configuration or to intercompare doses from these two system configurations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。