The root cause analysis on failed patient-specific measurements of pencil beam scanning protons using a 2D detection array with finite size ionization chambers

使用带有有限尺寸电离室的二维检测阵列对笔形束扫描质子进行患者特异性测量失败的根本原因分析

阅读:1

Abstract

The aim of this report is to present the root cause analysis on failed patient-specific quality assurance (QA) measurements of pencil beam scanning (PBS) protons; referred to as PBS-QA measurement. A criterion to fail a PBS-QA measurement is having a <95% passing rate in a 3.0%-3.0 mm gamma index analysis. Clinically, we use a two-dimensional (2D) gamma index analysis to obtain the passing rate. The IBA MatriXX PT 2D detection array with finite size ionization chamber was utilized. A total of 2488 measurements performed in our PBS beamline were cataloged. The percentage of measurements for the sites of head/neck, breast, prostate, and other are 53.3%, 22.7%, 10.5%, and 13.5%, respectively. The measurements with a passing rate of 100 to >94%, 94 to >88%, and <88% were 93.6%, 5.6%, and 0.8%, respectively. The percentage of failed measurements with a <95% passing rate was 10.9%. After removed the user errors of either re-measurement or re-analysis, 8.1% became acceptable. We observed a feature of >3% per mm dose gradient with respect to depth on the failed measurements. We utilized a 2D/three-dimensional (3D) gamma index analysis toolkit to investigate the effect of depth dose gradient. By utilizing this 3D toolkit, 43.1% of the failed measurements were improved. A feature among measurements that remained sub-optimal after re-analysis was a sharp >3% per mm lateral dose gradient that may not be well handled using the detector size of 5.0 mm in-diameter. An analysis of the sampling of finite size detectors using one-dimensional (1D) error function showed a large dose deviation at locations of low-dose areas between two high-dose plateaus. User error, large depth dose gradient, and the effect of detector size are identified as root causes. With the mitigation of the root causes, the goals of patient-specific QA, specifically detecting actual deviation of beam delivery or identifying limitations of the dose calculation algorithm of the treatment planning system, can be directly related to failure of the PBS-QA measurements.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。