Automatic couch position calculation using eclipse scripting for external beam radiotherapy

使用Eclipse脚本自动计算外照射放射治疗的治疗床位置

阅读:1

Abstract

PURPOSE: The treatment couch position of a patient in external beam radiation therapy (EBRT) is usually acquired during initial treatment setup. This procedure has shown potential failure modes leading to near misses and adverse events in radiation treatment. This study aims to develop a method to automatically determine the couch position before setting up a patient for initial treatment. METHODS: The Qfix couch-tops (kVue and DoseMax) have embedded reference marks (BBs) indicating its index levels and couch centerline. With the ESAPI, a C# script was programmed to automatically find the couch-top and embedded BBs in the planning CT and derive the treatment couch position according to treatment isocenter of a plan. Couch positions of EBRT plans with the kVue couch-top and SBRT plans using the DoseMax were calculated using the script. The calculation was evaluated by comparing calculated positions with couch coordinates captured during the initial treatment setup after image guidance. The calculations were further compared with daily treatment couch positions post image-guided adjustment for each treatment fraction. RESULTS: For plans using the kVue couch-top for various treatment sites, the median (5-95 percentiles) differences between calculated and captured couch positions were 0.1 (-0.2 - 0.9), 0.5 (-1.1-2.0), 0.10 (-1.3-1.3) cm in the vertical, longitudinal, and lateral direction respectively. For the DoseMax couch-top, the median differences were 0.1 (-0.2-0.7), 0.2 (-0.3-1.1), and 0.2 (-0.7-0.9) cm in respective direction. The calculated positions were within 1 and 2 cm from the mean fraction positions for 95% patients on DoseMax and kVue couch-top respectively. CONCLUSIONS: A method that automatically and accurately calculates treatment couch position from simulation CT was implemented in Varian Eclipse for Qfix couch-tops. This technique increases the efficiency of patient setup and enhances patient safety by reducing the risks of positioning errors.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。