Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator

对配备 WEHR 和 MEHRS 准直器的 GE NM/CT 870 CZT 临床 SPECT 系统进行实验评估

阅读:1

Abstract

PURPOSE: A high-energy-resolution whole-body SPECT-CT device (NM/CT 870 CZT; C-SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium-energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C-SPECT, and a NaI scintillation detector-equipped Anger-type SPECT (A-SPECT) scanner, with a LEHR and LMEGP. METHODS: A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using (99m) Tc and (123) I. We also measured the system's sensitivity and high-contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high-contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction. RESULTS: The energy resolution with (99m) Tc was 5.6% in C-SPECT with WEHR and 9.9% in A-SPECT with LEHR. Using 123I, the results were 9.1% in C-SPECT with WEHR, 5.5% in C-SPECT with MEHRS, and 10.4% in A-SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C-SPECT performed better in SPECT acquisition. High-contrast resolution was improved in C-SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C-SPECT. CONCLUSION: C-SPECT demonstrates excellent energy resolution and improved high-contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C-SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high-contrast resolution.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。