On the MLC leaves alignment in the direction orthogonal to movement

在多叶准直器上,叶片排列方向与运动方向正交。

阅读:1

Abstract

The main focus of the recommended spatial accuracy tests for the multi-leaf collimators (MLC) is calibration of the leaf position along the movement direction and overall alignment to the radiation isocenter. No explicit attention was typically paid to the alignment of the leaves from the opposing banks in the direction orthogonal to movement. This paper is a case study demonstrating that verification of such alignment at the time of acceptance testing is prudent. The original standard MLC (SMLC) on an MRIdian MRI-guided linac (ViewRay Inc., Mountain View, CA, USA) was upgraded to a high-speed MLC (HSMLC), which is supposed to be mechanically identical to the SMLC except for the higher drive screw pitch. The results of the end-to-end IMRT tests demonstrated unacceptable dosimetric results exemplified by an average and maximum ion chamber (IC) point dose error in the high-dose low-gradient region of 2.5 ± 1.4% and 4.6%, respectively. Before the upgrade, those values were 0.3 ± 0.7% and 0.9%, respectively. An exhaustive analysis of possible failure modes eventually zeroed in on the average misalignment of about 1 mm in the Y (along the couch) direction between the right and left upper MLC banks. The MLC was replaced, reducing the Y-direction misalignment to 0.4 mm. As a result, the average and maximum IC dose-errors became acceptable 1.0 ± 0.7% and 1.6%, respectively. Simple film and/or chamber array tests during acceptance testing can easily detect Y-direction misalignments between opposing leaves banks measuring a fraction of a mm at isocenter. Left undetected, such misalignment can cause nontrivial dosimetric consequences.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。