Characterization of radiotherapy component impact on MR imaging quality for an MRgRT system

表征放射治疗组件对磁共振引导放射治疗(MRgRT)系统磁共振成像质量的影响

阅读:1

Abstract

Radiotherapy components of an magnetic resonnace-guided radiotherapy (MRgRT) system can alter the magnetic fields, causing spatial distortion and image deformation, altering imaging and radiation isocenter coincidence and the accuracy of dose calculations. This work presents a characterization of radiotherapy component impact on MR imaging quality in terms of imaging isocenter variation and spatial integrity changes on a 0.35T MRgRT system, pre- and postupgrade of the system. The impact of gantry position, MLC field size, and treatment table power state on imaging isocenter and spatial integrity were investigated. A spatial integrity phantom was used for all tests. Images were acquired for gantry angles 0-330° at 30° increments to assess the impact of gantry position. For MLC and table power state tests all images were acquired at the home gantry position (330°). MLC field sizes ranged from 1.66 to 27.4 cm edge length square fields. Imaging isocenter shift caused by gantry position was reduced from 1.7 mm at gantry 150° preupgrade to 0.9 mm at gantry 120° postupgrade. Maximum spatial integrity errors were 0.5 mm or less pre- and postupgrade for all gantry angles, MLC field sizes, and treatment table power states. However, when the treatment table was powered on, there was significant reduction in SNR. This study showed that gantry position can impact imaging isocenter, but spatial integrity errors were not dependent on gantry position, MLC field size, or treatment table power state. Significant isocenter variation, while reduced postupgrade, is cause for further investigation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。