Technical Note: Extended field-of-view (FOV) MRI distortion determination through multi-positional phantom imaging

技术说明:通过多位置体模成像确定扩展视野(FOV)MRI畸变

阅读:1

Abstract

Comprehensive characterization of geometric distortions for MRI simulators and MRI-guided treatment delivery systems is typically performed with large phantoms that are costly and unwieldy to handle. Here we propose an easily implementable methodology for MR distortion determination of the entire imaging space of the scanner through the use of a compact commercially available distortion phantom. The MagphanRT phantom was scanned at several locations within a MR scanner. From each scan, an approximate location of the phantom was determined from a subset of the fiducial spheres. The fiducial displacements were determined, and a displacement field was fitted to the displacement data using the entire multi-scan data set. An orthogonal polynomial expansion fitting function was used that had been augmented to include independent rigid-body transformations for each scan. The rigid-body portions of the displacement field were thereafter discarded, and the resultant fit then represented the distortion field. Multi-positional scans of the phantom were used successfully to determine the distortion field with extended coverage. A single scan of the phantom covered 20 cm in its smallest dimension. By stitching together overlapping scans we extended the distortion measurements to 30 cm. No information about the absolute location or orientation of each scan was required. The method, termed the Multi-Scan Expansion (MSE) method, can be easily applied for larger field-of-views (FOVs) by using a combination of larger phantom displacements and more scans. The implementation of the MSE method allows for distortion determination beyond the physical limitations of the phantom. The method is scalable to the user's needs and does not require any specialized equipment. This approach could open up for easier determination of the distortion magnitude at distances further from the scanner's isocenter. This is especially important in the newly proposed methodologies of MR-only simulation in RT and in adaptive replanning in MR linac systems.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。