TomoMQA: Automated analysis program for MVCT quality assurance of helical tomotherapy

TomoMQA:用于螺旋断层放射治疗MVCT质量保证的自动化分析程序

阅读:1

Abstract

PURPOSE: In this study, we developed a simple but useful computer program, called TomoMQA, to offer an automated quality assurance for mega-voltage computed tomography (MVCT) images generated via helical tomotherapy. METHODS: TomoMQA is written in MATLAB and contains three steps for analysis: (a) open the DICOM dataset folder generated via helical tomotherapy (i.e., TomoTherapy(®) and Radixact™), (b) call the baseline data for the consistency test and click the "Analysis" button (or click the "Analysis" button without the baseline data and export the results as the baseline data), and (c) print an analyzed report. The overall procedure for the QA analysis included in TomoMQA is referred from the TG-148 recommendation. Here, the tolerances for MVCT QA were implemented from TG-148 recommended values as default; however, it can be modified by a user manually. RESULTS: To test the performance of the TomoMQA program, 15 MVCTs were prepared from five helical tomotherapy machines (1 of TomoTherapy(®) HD, 2 of TomoTherapy(®) HDA, and 2 of Radixact™) in 3 months and the QA procedures were performed using TomoMQA. From our results, the evaluation revealed that the developed program can successfully perform the MVCT QA analysis irrespective of the type of helical tomotherapy equipment. CONCLUSION: We successfully developed a new automated analysis program for MVCT QA of a helical tomotherapy platform, called TomoMQA. The developed program will be made freely downloadable from the TomoMQA-dedicated website.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。