A novel and effective method for validation and measurement of output factors for Leksell Gamma Knife® Icon™ using TRS 483 protocol

一种采用TRS 483协议验证和测量Leksell Gamma Knife® Icon™输出因子的新颖有效方法

阅读:1

Abstract

The objective of this work was to identify the exact location of the effective point of measurement (EPM) of four different active detectors to compare the relative collimator output factors (ROF) of Leksell Gamma Knife (LGK) according to IAEA TRS-483 recommendations. ROF was measured at the center of the spherical LGK-Solid Water (LGK-SW) Phantom for three (4-, 8-, and 16-mm in diameter) collimators using four (PTW-TN60008, PTW-TN60016, PTW-TN60017, and PTW-60019 diode/diamond) detectors. Since diode detectors have a much smaller sensitive volume than the PTW-31010 ion chamber used for reference dosimetry, its EPM might not be at the center of the phantom, or (100, 100, 100) of the Leksell Coordinate System, particularly in the z-direction. Hence for each diode detector, a CBCT image was acquired after it was inserted into the phantom, from which the z-Leksell coordinate of EPM was determined. Relative collimator output factors was then measured by focusing GK beams on the determined EPM of each diode. Measured ROFs were compared with the vendor-provided values in GK treatment planning system. For validation, a plan was generated to measure the output of 4-mm collimator for PTW-TN60017 at various couch locations along the z-axis. For PTW-TN60008, the percentage variations were 0.6 ± 0.4%, and -0.8 ± 0.2% for 4 and 8-mm collimators, respectively. For PTW-TN60016, the percentage variations were 0.8 ± 0.0%, and 0.2 ± 0.1%, respectively. The percentage variations were -3.3 ± 0.0% and -0.9 ± 0.1%, respectively, for PTW-TN60017, and -0.5 ± 0.0% and -0.8 ± 0.2%, respectively, for PTW-TN60019. Center of the measured profile for PTW-TN60017 was only 0.1 mm different from that identified using the CBCT. In conclusion, we have developed a simple and effective method to determine the EPMs of diode detectors when inserted into the existing LGK-SW phantom. With the acquired positional information and using TRS-483 protocol, good agreements were obtained between the measured ROFs and manufacturer recommended values.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。