Use of a plastic scintillator detector for patient-specific quality assurance of VMAT SRS

使用塑料闪烁体探测器进行VMAT SRS的患者特异性质量保证

阅读:1

Abstract

PURPOSE: To evaluate a scintillator detector for patient-specific quality assurance of VMAT radiosurgery plans. METHODS: The detector was comprised of a 1 mm diameter, 1 mm high scintillator coupled to an acrylic optical fiber. Sixty VMAT SRS plans for treatment of single targets having sizes ranging from 3 mm to 30.2 mm equivalent diameter (median 16.3 mm) were selected. The plans were delivered to a 20 cm × 20 cm x 15 cm water equivalent plastic phantom having either the scintillator detector or radiochromic film at the center. Calibration films were obtained for each measurement session. The films were scanned and converted to dose using a 3-channel technique. RESULTS: The mean difference between scintillator and film was -0.45% (95% confidence interval -0.1% to 0.8%). For target equivalent diameter smaller than the median, the mean difference was 1.1% (95% confidence interval 0.5% to 1.7%). For targets larger than the median, the mean difference was -0.2% (95% confidence interval -0.7% to 0.1%). CONCLUSIONS: The scintillator detector response is independent of target size for targets as small as 3 mm and is well-suited for patient-specific quality assurance of VMAT SRS plans. Further work is needed to evaluate the accuracy for VMAT plans that treat multiple targets using a single isocenter.