A simulation-based method for evaluating geometric tests of a linac c-arm in quality control in radiotherapy

一种基于模拟的放射治疗质量控制中直线加速器C臂几何测试评估方法

阅读:1

Abstract

PURPOSE: Assessment of the accuracy of geometric tests of a linac used in external beam therapy is crucial for ensuring precise dose delivery. In this paper, a new simulation-based method for assessing accuracy of such geometric tests is proposed and evaluated on a set of testing procedures. METHODS: Linac geometry testing methods used in this study are based on an established design of a two-module phantom. Electronic portal imaging device (EPID) images of fiducial balls contained in these modules can be used to automatically reconstruct linac geometry. The projection of the phantom modules fiducial balls onto the EPID detector plane is simulated for assumed nominal geometry of a linac. Then, random errors are added to the coordinates of the projections of the centers of the fiducial balls and the linac geometry is reconstructed from these data. RESULTS: Reconstruction is performed for a set of geometric test designs and it is shown how the dispersion of the reconstructed values of geometric parameters depends on the design of a geometric test. Assuming realistic accuracy of EPID image analysis, it is shown that for selected testing plans the reconstruction accuracy of geometric parameters can be significantly better than commonly used action thresholds for these parameters. CONCLUSIONS: Proposed solution has the potential to improve geometric testing design and practice. It is an important part of a fully automated geometric testing solution.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。