Applying three different methods of measuring CTDI(free air) to the extended CTDI formalism for wide-beam scanners (IEC 60601-2-44): A comparative study

将三种不同的CTDI(自由空气)测量方法应用于宽束扫描仪的扩展CTDI公式(IEC 60601-2-44):一项比较研究

阅读:1

Abstract

PURPOSE: The weighted CT dose index (CTDI(w) ) has been extended for a nominal total collimation width (nT) greater than 40 mm and relies on measurements of CTDIfreeair. The purpose of this work was to compare three methods of measuring CTDIfreeair and subsequent calculations of CTDI(w) to investigate their clinical appropriateness. METHODS: The CTDIfreeair, for multiple nTs up to 160 mm, was calculated from (1) high-resolution air kerma profiles from a step-and-shoot translation of a liquid ionization chamber (LIC) (considered to be a dosimetric reference), (2) pencil ionization chamber (PIC) measurements at multiple contiguous positions, and (3) air kerma profiles obtained through the continuous translation of a solid-state detector. The resulting CTDIfreeair was used to calculate the CTDI(w) , per the extended formalism, and compared. RESULTS: The LIC indicated that a 40 mm nT should not be excluded from the extension of the CTDI(w) formalism. The solid-state detector differed by as much as 8% compared to the LIC. The PIC was the most straightforward method and gave equivalent results to the LIC. CONCLUSIONS: The CTDI(w) calculated with the latest CTDI formalism will differ most for 160 mm nTs (e.g., whole-organ perfusion or coronary CT angiography) compared to the previous CTDI formalism. Inaccuracies in the measurement of CTDIfreeair will subsequently manifest themselves as erroneous calculations of the CTDI(w) , for nTs greater than 40 mm, with the latest CTDI formalism. The PIC was found to be the most clinically feasible method and was validated against the LIC.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。