Reproducibility of radioactive iodine uptake (RAIU) measurements

放射性碘摄取率(RAIU)测量的可重复性

阅读:1

Abstract

BACKGROUND: Measurement of radioactive iodine uptake (RAIU) is an important aspect of the assessment and treatment of patients with hyperthyroidism. Its uncertainty affects how much of a true change in RAIU can be detected as well as appropriateness of the therapy dosage upon which it is based. In this study, a method of estimating the reproducibility and least significant change (LSC) values for RAIU measurements, and the implications of the values observed are discussed, with emphasis on application to quality assurance initiatives. METHODS: We prospectively studied 36 consecutive patients referred for RAIU measurements. Twenty-four hours after oral administration of 370 kBq of (131) I-NaI in capsule form, RAIU measurements were obtained in duplicate using a thyroid probe uptake system. Assessment of reproducibility was performed using root-mean-square standard deviation. RESULTS: Average difference between duplicated RAIU measurements in our study cohort was -0.1 ± 1.6% and ranged from -4.8% to 3.1%. Reproducibility of probe-based RAIU measurement was calculated to be 1.1% and 95% LSC was 3.2%. CONCLUSION: In our clinic, probe-based RAIU is a reproducible and relatively precise measurement. Using the method we have outlined, each institution can perform reproducibility assessment and compute the LSC of RAIU measurements based on its own staff, iodine isotope, equipment, protocols, and patient population. These values are useful in the assessment of serial change in RAIU, and as more experience is accumulated, can serve as benchmarks to be used in quality assurance initiatives.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。