Relationship between the radiation doses at nonenhanced CT studies using different tube voltages and automatic tube current modulation during anthropomorphic phantoms of young children

在对幼儿进行人体模型检查时,采用不同管电压和自动管电流调节的非增强CT检查中辐射剂量之间的关系

阅读:2

Abstract

To compare the radiation dose and image noise of nonenhanced CT scans performed at 80, 100, and 120 kVp with tube current modulation (TCM) we used anthropomorphic phantoms of newborn, 1-year-old, and 5-year-old children. The noise index was set at 12. The image noise in the center of the phantoms at the level of the chest and abdomen was measured within a circumscribed region of interest. We measured the doses in individual tissues or organs with radio-photoluminescence glass dosimeters for each phantom. Various tissues or organs were assigned and the radiation dose was calculated based on the international commission on radiological protection definition. With TCM the respective radiation dose at tube voltages of 80, 100, and 120 was 29.71, 31.60, and 33.79 mGy for the newborn, 32.00, 36.79, and 39.48 mGy for the 1-year-old, and 32.78, 38.11, and 40.85 mGy for the 5-year-old phantom. There were no significant differences in the radiation dose among the tube voltages and phantoms (P > 0.05). Our comparison of the radiation dose using anthropomorphic phantoms of young children showed that the radiation dose of nonenhanced CT performed at different tube voltages with TCM was not significantly different.