The impact of a high-definition multileaf collimator for spine SBRT

高分辨率多叶准直器对脊柱立体定向放射治疗的影响

阅读:1

Abstract

PURPOSE: Advanced radiotherapy delivery systems designed for high-dose, high-precision treatments often come equipped with high-definition multi-leaf collimators (HD-MLC) aimed at more finely shaping radiation dose to the target. In this work, we study the effect of a high definition MLC on spine stereotactic body radiation therapy (SBRT) treatment plan quality and plan deliverability. METHODS AND MATERIALS: Seventeen spine SBRT cases were planned with VMAT using a standard definition MLC (M120), HD-MLC, and HD-MLC with an added objective to reduce monitor units (MU). M120 plans were converted into plans deliverable on an HD-MLC using in-house software. Plan quality and plan deliverability as measured by portal dosimetry were compared among the three types of plans. RESULTS: Only minor differences were noted in plan quality between the M120 and HD-MLC plans. Plans generated with the HD-MLC tended to have better spinal cord sparing (3% reduction in maximum cord dose). HD-MLC plans on average had 12% more MU and 55% greater modulation complexity as defined by an in-house metric. HD-MLC plans also had significantly degraded deliverability. Of the VMAT arcs measured, 94% had lower gamma passing metrics when using the HD-MLC. CONCLUSION: Modest improvements in plan quality were noted when switching from M120 to HD-MLC at the expense of significantly less accurate deliverability in some cases.