Validating k(Q) =1.0 assumption in TG51 with PTW 30013 farmer chamber for Varian TrueBeam's 2.5 MV imaging beam

使用PTW 30013 Farmer电离室在TG51中验证Varian TrueBeam 2.5 MV成像束的k(Q) = 1.0假设

阅读:1

Abstract

AAPM Report 142 recommends and the State of Ohio requires that the imaging dose be quantified in radiotherapy applications. Using the TG51 dose calibration protocol for MV Imaging dose measurement requires knowledge of the k(Q) parameter for the beam quality and the ionization chamber type under investigation. The %dd(10)(x) of the Varian TrueBeam 2.5 MV imaging beam falls outside the range of the available data for the calculation of the k(Q) value. Due to the similarities of the 2.5 MV imaging beam and the (60) Co beam, we and others made the assumption that k(Q) = 1.0 in TG51 calculations. In this study, we used the TG21 and TG51 calibration protocols in conjunction to validate that k(Q) = 1.0 for the 2.5 MV imaging beam using a PTW 30013 farmer chamber. Standard measurements for TG51 absolute dosimetry QA were performed at 100 cm SSD, 10 cm depth, 10 × 10 field size, delivering 100 Monitor Units to a waterproof Farmer Chamber (PTW TN30013) for both 2.5 and 6 MV. Both the TG21 and TG51 formalisms were used to calculate the dose to water per MU at d(max) (D(w) /MU) for the 6 MV beam. The calculated outputs were 1.0005 and 1.0004 cGy/MU respectively. The TG21 formalism was then used to calculate (D(w) /MU) for the 2.5 MV imaging beam. This value was then used in the TG51 formalism to find k(Q) for the 2.5 MV imaging beam. A k(Q) value of 1.00 ± 0.01 was calculated for 2.5 MV using this method.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。