Use of reduced dose rate when treating moving tumors using dynamic IMRT

使用动态调强放射治疗(IMRT)治疗移动肿瘤时,应采用降低剂量率的方法。

阅读:1

Abstract

The purpose was to evaluate the effect of dose rate on discrepancies between expected and delivered dose caused by the interplay effect. Fifteen separate dynamic IMRT plans and five hybrid IMRT plans were created for five patients (three IMRT plans and one hybrid IMRT plan per patient). The impact of motion on the delivered dose was evaluated experimentally for each treatment field for different dose rates (200 and 400 MU/min), and for a range of target amplitudes and periods. The maximum dose discrepancy for dynamic IMRT fields was 18.5% and 10.3% for dose rates of 400 and 200 MU/min, respectively. The maximum dose discrepancy was larger than this for hybrid plans, but the results were similar when weighted by the contribution of the IMRT fields. The percentage of fields for which 98% of the target never experienced a 5% or 10% dose discrepancy increased when the dose rate was reduced from 400 MU/min to 200 MU/min. For amplitudes up to 2 cm, reducing the dose rate to 200 MU/min is effective in keeping daily dose discrepancies for each field within 10%.