Characterization of Nyquist ghost in EPI-fMRI acquisition sequences implemented on two clinical 1.5 T MR scanner systems: effect of readout bandwidth and echo spacing

在两台临床1.5T磁共振扫描仪系统上实现的EPI-fMRI采集序列中奈奎斯特鬼像的特征分析:读出带宽和回波间隔的影响

阅读:1

Abstract

In EPI-fMRI acquisitions, various readout bandwidth (BW) values are used as a function of gradients' characteristics of the MR scanner system. Echo spacing (ES) is another fundamental parameter of EPI-fMRI sequences, but the employed ES value is not usually reported in fMRI studies. Nyquist ghost is a typical EPI artifact that can degrade the overall quality of fMRI time series. In this work, the authors assessed the basic effect of BW and ES for two clinical 1.5 T MR scanner systems (scanner-A, scanner-B) on Nyquist ghost of gradient-echo EPI-fMRI sequences. BW range was: scanner-A, 1953-3906 Hz/pixel; scanner-B, 1220-2894 Hz/pixel. ES range was: scanner-A, scanner-B: 0.75-1.33 ms. The ghost-to-signal ratio of time series acquisition (GSRts) and drift of ghost-to-signal ratio (DRGSR) were measured in a water phantom. For both scanner-A (93% of variation) and scanner-B (102% of variation) the mean GSRts significantly increased with increasing BW. GSRts values of scanner-A did not significantly depended on ES. On the other hand, GSRts values of scanner-B significantly varied with ES, showing a downward trend (81% of variation) with increasing ES. In addition, a GSRts spike point at ES = 1.05 ms indicating a potential resonant effect was revealed. For both scanners, no significant effect of ES on DRGSR was revealed. DRGSR values of scanner-B did not significantly vary with BW, whereas DRGSR values of scanner-A significantly depended on BW showing an upward trend from negative to positive values with increasing BW. GSRts and DRGSR can significantly vary with BW and ES, and the specific pattern of variation may depend on gradients performances, EPI sequence calibrations and functional design of radiofrequency coil. Thus, each MR scanner system should be separately characterized. In general, the employment of low BW values seems to reduce the intensity and temporal variation of Nyquist ghost in EPI-fMRI time series. On the other hand, the use of minimum ES value might not be entirely advantageous when the MR scanner is characterized by gradients with low performances and suboptimal EPI sequence calibration.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。