Initial evaluation of a four-dimensional computed tomography system using a programmable motor

使用可编程电机的四维计算机断层扫描系统初步评估

阅读:1

Abstract

A dynamic lung tumor phantom was used to investigate the geometric reconstruction accuracy of a commercial four-dimensional computed tomography (4D-CT) system. A ball filled with resin, embedded in a cork cube, was placed on a moving platform. Various realistic antero-posterior (AP) motions were programmed to reproduce the respiratory motion of a lung tumor. Several three-dimensional (3D) CT and 4D-CT images of this moving object were acquired and compared using different acquisition parameters. Apparent volume and diameter of the ball were measured and compared to the real values. The position of two points (the AP limits of the ball) during the motion in the coordinate system of the CT scanner were also compared with the expected values. Volume error was shown to increase with object speed. However, although the volume error was associated with intraslice artifacts, it did not reflect large interslice inconstancies in object position and should not be used as an indicator of image accuracy. The 3D-CT gave a random position of the tumor along the phantom excursion; accuracy in the assessment of position by 4D-CT ranged from 0.4 mm to 2.6 mm during extreme phases of breathing. We used average projection (AVE) and maximum intensity projection (MIP) algorithms available on the commercial software to create internal target volumes (ITVs) by merging gross tumor volume (GTV) images at various respiratory phases. The ITVs were compared to a theoretical value computed from the programmed ball excursion. The ITVs created from the MIP algorithm were closer to the theoretical value (within 12%) than were those created from the AVE algorithm (within 40%).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。