Radiographic techniques in screen-film mammography

乳腺X线摄影中的放射成像技术

阅读:1

Abstract

The objectives of this study were to document imaging physics parameters associated with mammography physics surveys, and investigate how the choice of tube potential affects average glandular dose (AGD) and x-ray exposure time. Data from 60 mammography units were obtained pertaining to representative values of mAs, exposure time, half value layer, AGD and film density when acquiring phantom images. The survey of clinical systems showed that for a normal sized breast as represented by the mammography accreditation phantom, 60% of these units were operated at 25 kVp, and 33% at 26 kVp. Median exposure times were 1.14 s at 25 kVp and 0.73 s at 26 kVp. The median AGD was 1.62 mGy at 25 kVp and 1.51 mGy at 26 kVp. As expected, the choice of x-ray tube potential did not significantly affect the median film density value of 1.5. Five clinical systems, all from different vendors, had measurements performed of the AGD and x-ray exposure time as a function of x-ray tube potential at a constant film density. For a typical clinical x-ray unit, increasing the x-ray tube potential from 25 to 28 kVp reduced the exposure time by 50%, and reduced the AGD by 26%.